TESTEC Elektronik GmbH
TESTEC TT-SI 7005 Active Differential Probe | 60MHz, ±35V, x1/x5/x10 Attenuation
TESTEC TT-SI 7005 Active Differential Probe | 60MHz, ±35V, x1/x5/x10 Attenuation

Key Features
Overview
The TESTEC TT-SI 7005 is an active differential probe with switchable x1 / x5 / x10 attenuation, bandwidth up to 60 MHz and a maximum differential input of ±35 V. It lets a ground-referenced oscilloscope measure between two points where neither is at earth, such as current-sense shunts, differential sensor outputs and low-voltage bus signals. It runs from a 9 V adapter and has a 1 m BNC output cable.
Downloads
User Manuals
- Details
TESTEC TT-SI 7005 Differential Probe
The TESTEC TT-SI 7005 is an active differential probe with switchable x1 / x5 / x10 attenuation, bandwidth up to 60 MHz and a maximum differential input of ±35 V. It lets a ground-referenced oscilloscope measure between two points where neither is at earth, such as current-sense shunts, differential sensor outputs and low-voltage bus signals. It runs from a 9 V adapter and has a 1 m BNC output cable.
Floating Measurements on a Grounded Oscilloscope
A differential probe subtracts one input from the other and sends the difference to the oscilloscope as a ground-referenced signal. That removes the need to float the oscilloscope or lift its earth, and lets you look at a high-side node and a low-side node on the same instrument. Common-mode rejection is 80 dB at 60 Hz, 60 dB at 100 Hz and 50 dB at 1 MHz.
Ranges and Bandwidth
Each attenuation setting has its own range: ±3.5 V at x1 and ±18 V at x5 and ±35 V at x10. Bandwidth is 15 MHz at x1 and 50 MHz at x5 and 60 MHz at x10, with rise times of 23 ns at x1, 7 ns at x5 and 5.8 ns at x10. Differential input impedance is 1 MΩ ∥ 2.5 pF; each side to ground is 0.5 MΩ ∥ 6 pF. The output swings ±4 V into a 1 MΩ oscilloscope input, with typical offset of ±5 mV and noise of 2 mVrms. DC accuracy is ±2%. Power and over-range LEDs show the probe's state.
Rating to Ground
TESTEC builds the probe to IEC/EN 61010-031:2015 CAT III, pollution degree 2. TESTEC's TT-SI 7005 manual gives an absolute maximum of 12 Vrms from each input to ground. The oscilloscope must be earthed.
Limitations
- Each input is limited to 12 Vrms to ground.
- Needs mains power from the 9 V adapter; it is not battery powered. The TT-SX 9001 runs from AA cells or USB-C.
- Needs a 1 MΩ oscilloscope input for the specified ±4 V output swing.
- Low voltage only: ±35 V differential at most. For higher voltages choose the TT-SI 7002 or TT-SX 9001.
TT-SI 7005 Compared with Other TESTEC Differential Probes
| Model | Bandwidth | Attenuation | Maximum differential voltage | Power | Price |
|---|---|---|---|---|---|
| TT-SI 7002 | 25 MHz | x20 / x50 / x200 | ±700 V | 9 V adapter | £300 |
| TT-SI 7005 (this product) | 60 MHz | x1 / x5 / x10 | ±35 V | 9 V adapter | £425 |
| TT-SX 9001 | 30 MHz | x10 / x100 | ±700 V | 4 × AA or USB-C | £555 |
| TT-SX 9002 | 30 MHz | x20 / x200 | ±1400 V | 4 × AA or USB-C | £555 |
| TT-SI 8071 | 70 MHz | x10 / x100 | ±700 V | USB | £690 |
| TT-SX 200 | 200 MHz | 10:1 | ±20 V | USB | £1,300 |
Typical Applications
Measuring the voltage across a current-sense shunt that is not at ground. Probing differential sensor and audio outputs. Checking low-voltage differential signals such as CAN and RS-485 at the physical layer.
The TT-SI 7005 suits embedded and analogue engineers who need clean floating measurements of small signals. See all TESTEC differential probes.
Order the TESTEC TT-SI 7005 from £425 ex VAT. Tell us the voltage to ground in your circuit and we will confirm the right differential probe before you order.
Last reviewed: 23 November 2026
Sources (4 TESTEC documents and pages)
- TESTEC TT-SI 7002 and TT-SI 7005 datasheet (PDF)
- TESTEC differential probes product page (testec.de)
- TESTEC differential probes comparison chart (PDF)
- TESTEC TT-SI 7005 instruction manual (in the downloads on this page)
| General | |
|---|---|
| Order number | 15176 |
| Probe type | Active differential probe |
| Attenuation | x1 / x5 / x10 |
| Bandwidth | 15 MHz at x1; 50 MHz at x5; 60 MHz at x10 |
| Rise time | 23 ns at x1, 7 ns at x5 and 5.8 ns at x10 |
| Input | |
| Input impedance | Differential: 1 MΩ ∥ 2.5 pF; Each side to ground: 0.5 MΩ ∥ 6 pF |
| Maximum differential voltage | DC + AC peak: ±3.5 V at x1; ±18 V at x5; ±35 V at x10 |
| Maximum voltage to ground | 12 Vrms |
| Accuracy and noise | |
| DC accuracy | ±2% |
| Noise | Typical: 2 mVrms |
| CMRR | 80 dB at 60 Hz; 60 dB at 100 Hz; 50 dB at 1 MHz |
| Offset | Typical: ±5 mV |
| Output swing | ±4 V into 1 MΩ |
| Connection and power | |
| Power supply | 9 V adapter |
| Cable length | BNC cable: 100 cm |
| Input leads | 55 cm each |
| Safety and environment | |
| Safety standard | IEC/EN 61010-031:2015 CAT III, pollution degree 2 |
| Operating temperature | 0 °C to 50 °C |
| Humidity | 10% to 85% RH |
| Mechanical | |
| Dimensions | 195 × 55 × 30 mm |
| Weight | 250 g |
| Ordering information | |
| Part number (SKU) | 15176 |
| Manufacturer | TESTEC Elektronik GmbH |
| EAN | 5055383624683 |
| Warranty | 12 months |
| Shipping weight | 1.5 kg |
- 1 ×TT-SI 7005 differential probe
- 2 ×GR2 grabbers
- 2 ×TL7 4 mm silicone test leads
- BN7 BNC cable (100 cm)
- NT7 9 V adapter (EU plug)
- Manual
Questions and Answers
-
What Voltage Can the TT-SI 7005 Measure?
Up to ±35 V differential (DC + AC peak) at x10, ±18 V at x5 and ±3.5 V at x1. Each input is limited to 12 Vrms to ground, per TESTEC's manual.
-
What Bandwidth Does It Have (TT-SI 7005)?
60 MHz at x10, 50 MHz at x5 and 15 MHz at x1.
-
Why Choose X1 Attenuation (TT-SI 7005)?
At x1 the full signal reaches the oscilloscope, which suits millivolt-level signals across shunts and sensors, at up to 15 MHz.
No questions match your search.