TESTEC Elektronik GmbH

TESTEC TT-LX 312 Passive Oscilloscope Probe

TESTEC TT-LX 312 Passive Oscilloscope Probe

Skip to product information
1 of 1

SKU: 10120

Regular price
£35.00 exc VAT
£42.00 inc VAT
Regular price Sale price
£35.00 exc VAT
£42.00 inc VAT
Sale Sold out
Shipping calculated at checkout.
If no stock is shown above, order now and we'll dispatch within 5 working days.

Payment methods

  • American Express
  • Visa
  • Mastercard
  • Maestro
  • PayPal
  • Apple Pay
  • Google Pay
  • Bank transfer
View full details

Key Features

150 MHz Bandwidth
2.3 ns Rise Time
x1 / x10 Switchable Attenuation
400 Vrms CAT II Maximum Input
10 MΩ, 13 pF Input Impedance
8 pF to 25 pF Compensation
±2% DC Accuracy

Overview

The TESTEC TT-LX 312 is a switchable x1/x10 passive oscilloscope probe: 150 MHz and 2.3 ns rise time at x10 with a 10 MΩ, 13 pF input, and 10 MHz with 1 MΩ, 47 pF at x1. It is rated 400 Vrms CAT II at x10 and 55 Vrms CAT II at x1, with a 1.2 m cable. One probe covers both small-signal checks and general digital and power work.

Downloads

User Manuals

  • TESTEC TT-LX 112 212 and 312 Oscilloscope Probes PDF824 KB · Aug 2005
    Details

The TESTEC TT-LX 312 is a switchable x1/x10 passive oscilloscope probe: 150 MHz and 2.3 ns rise time at x10 with a 10 MΩ, 13 pF input, and 10 MHz with 1 MΩ, 47 pF at x1. It is rated 400 Vrms CAT II at x10 and 55 Vrms CAT II at x1, with a 1.2 m cable. One probe covers both small-signal checks and general digital and power work.

Switch Between x1 and x10 on One Probe

At x1 the TT-LX 312 passes the full signal to the oscilloscope for millivolt-level ripple and noise, at up to 10 MHz. Switch to x10 and the same probe becomes a 150 MHz, 10 MΩ, 13 pF probe for logic edges and switching waveforms. Remember to set the oscilloscope's probe factor to match the switch position, or the vertical scale will be out by ten.

x10 Performance and Compensation

In x10 mode the rise time is 2.3 ns. The compensation trimmer in the BNC connector box covers oscilloscope inputs from 8 pF to 25 pF; set it with a 1 kHz square wave or the oscilloscope's calibrator for a flat-topped waveform, each time the probe moves to another instrument or channel. TESTEC gives no compensation range for the x1 position.

Safety Rating

TESTEC rates the x10 setting at 400 Vrms CAT II with 1500 V transient overvoltage, and the x1 setting at 55 Vrms CAT II. The x10 rating derates with frequency along the curve in the manual. The probe is for grounded oscilloscopes with a 1 MΩ BNC input only, and the reference lead must only go to earth potential.

Limitations

  • x1 bandwidth is 10 MHz and the input is 47 pF. Use x10 for anything fast or high impedance.
  • x10 bandwidth is 150 MHz. For faster edges choose the TT-MX 312 (250 MHz) or TT-HX 312 (350 MHz), which keep the x1/x10 switch.
  • Ground-referenced only. Floating measurements need an active differential probe.

TT-LX 312 Compared with Other TESTEC 5 mm Probes

Model Attenuation Bandwidth Rise time Input (x10) Price
TT-LX 112 x1 25 MHz 14 ns 1 MΩ, 42 pF (x1) £30
TT-LX 212 x10 150 MHz 2.3 ns 10 MΩ, 13 pF £35
TT-LX 312 (this product) x1 / x10 10 MHz / 150 MHz 35 ns / 2.3 ns 10 MΩ, 13 pF £35
TT-MX 212 x10 250 MHz 1.4 ns 10 MΩ, 13 pF £40
TT-MX 312 x1 / x10 10 MHz / 250 MHz 35 ns / 1.4 ns 10 MΩ, 13 pF £40
TT-HX 212 x10 350 MHz 1 ns 10 MΩ, 13 pF £60
TT-HX 312 x1 / x10 10 MHz / 350 MHz 35 ns / 1 ns 10 MΩ, 13 pF £60

Typical Applications

Measuring rail ripple at x1, then switching to x10 to check the regulator's switch node. Fault-finding on mixed analogue and digital boards with one probe per channel. Teaching and lab benches where one probe type must cover every exercise.

The TT-LX 312 suits engineers and labs that want one general-purpose probe per channel on oscilloscopes up to about 150 MHz. Browse all TESTEC passive oscilloscope probes.

Order the TESTEC TT-LX 312 from £35 ex VAT. Equipping a teaching lab or test area? Ask us for a formal quotation.

Last reviewed: 23 November 2026

Sources (2 TESTEC documents and pages)
General
Order number10120
Probe typePassive oscilloscope probe
Attenuationx1 / x10, switchable
Bandwidthx1: 10 MHz; x10: 150 MHz
Rise timex1: 35 ns; x10: 2.3 ns
Input
Input impedancex1: 1 MΩ, 47 pF; x10: 10 MΩ, 13 pF
Maximum input voltagex1: 55 Vrms CAT II; x10: 400 Vrms CAT II, 1500 V transient overvoltage (x10 derating with frequency)
Compensation range8 pF to 25 pF (x10)
Accuracy and noise
DC accuracy±2%
Connection and power
Cable length1.2 m
Safety and environment
Safety standardEN 61010-031
Pollution degree2
Operating temperature0 °C to 50 °C
Humidity5% to 95% RH (10 °C to 30 °C); 5% to 50% RH above 40 °C
Operating altitudeUp to 3000 m
Ordering information
Part number (SKU)10120
ManufacturerTESTEC Elektronik GmbH
EAN5055383624034
Warranty12 months
Shipping weight0.5 kg
  • 1 × TESTEC TT-LX 312 Passive Oscilloscope Probe

Contact us to confirm the full contents of the box.

Engineer Q&A

Questions and Answers

3 questions
  • What Is the Bandwidth at X1 and X10 (TT-LX 312)?

    10 MHz at x1 and 150 MHz at x10, with rise times of 35 ns and 2.3 ns.

    Answered by John LeggUpdated 23 Nov 2026
  • What Voltage Can the TT-LX 312 Measure?

    55 Vrms CAT II at x1 and 400 Vrms CAT II at x10, with 1500 V transient overvoltage at x10. Both derate with frequency.

    Answered by John LeggUpdated 23 Nov 2026
  • How Do I Compensate the TT-LX 312?

    In the x10 position, apply a 1 kHz square wave or the oscilloscope's calibrator output and adjust the trimmer in the BNC box for a flat-topped waveform. The range is 8 pF to 25 pF.

    Answered by John LeggUpdated 23 Nov 2026

Can't find your answer? Ask our engineers.

Ask a Question

Ask Our Engineers About This Product

Your question goes to the Debug Store technical team. We answer by email, and useful questions may be added to this page without your name or email address.