TESTEC Elektronik GmbH
TESTEC TT-HX 212 350MHz Passive Oscilloscope Probe
TESTEC TT-HX 212 350MHz Passive Oscilloscope Probe

Key Features
Overview
The TESTEC TT-HX 212 is a fixed x10 passive oscilloscope probe with 350 MHz bandwidth, 1 ns rise time and a 10 MΩ, 13 pF input. It is rated 400 Vrms CAT II with 1500 V transient overvoltage and compensates oscilloscope inputs from 8 pF to 25 pF. It is the fastest of TESTEC's 5 mm probes, for 200 MHz to 350 MHz oscilloscopes.
The TESTEC TT-HX 212 is a fixed x10 passive oscilloscope probe with 350 MHz bandwidth, 1 ns rise time and a 10 MΩ, 13 pF input. It is rated 400 Vrms CAT II with 1500 V transient overvoltage and compensates oscilloscope inputs from 8 pF to 25 pF. It is the fastest of TESTEC's 5 mm probes, for 200 MHz to 350 MHz oscilloscopes.
350 MHz and 1 ns Rise Time
The HX series is TESTEC's top 5 mm passive range. At 350 MHz and 1 ns rise time it is the fastest probe in the LX, MX and HX families. A 1 ns rise time resolves sub-5 ns edges on fast logic, SPI flash interfaces and switching power stages, while 10 MΩ and 13 pF keep circuit loading low.
Compensation
The trimmer in the BNC connector box covers 8 pF to 25 pF. Use a 1 kHz square wave or the oscilloscope's calibrator and adjust for a flat-topped waveform whenever the probe moves to another channel or instrument.
Safety Rating
TESTEC rates the TT-HX 212 at 400 Vrms Measurement Category II with 1500 V transient overvoltage, derating with frequency. It is for grounded oscilloscopes with a 1 MΩ BNC input, and the reference lead must only go to earth potential.
Limitations
- Fixed x10; the TT-HX 312 adds a x1 position.
- For 500 MHz and a smaller 2.5 mm tip choose the TT-HF 612.
- Ground-referenced only. Floating measurements need an active differential probe.
TT-HX 212 Compared with Other TESTEC Passive Probes
| Model | Attenuation | Bandwidth | Rise time | Input (x10) | Price |
|---|---|---|---|---|---|
| TT-LX 112 | x1 | 25 MHz | 14 ns | 1 MΩ, 42 pF (x1) | £30 |
| TT-LX 212 | x10 | 150 MHz | 2.3 ns | 10 MΩ, 13 pF | £35 |
| TT-LX 312 | x1 / x10 | 10 MHz / 150 MHz | 35 ns / 2.3 ns | 10 MΩ, 13 pF | £35 |
| TT-MX 212 | x10 | 250 MHz | 1.4 ns | 10 MΩ, 13 pF | £40 |
| TT-MX 312 | x1 / x10 | 10 MHz / 250 MHz | 35 ns / 1.4 ns | 10 MΩ, 13 pF | £40 |
| TT-HX 212 (this product) | x10 | 350 MHz | 1 ns | 10 MΩ, 13 pF | £60 |
| TT-HX 312 | x1 / x10 | 10 MHz / 350 MHz | 35 ns / 1 ns | 10 MΩ, 13 pF | £60 |
| TT-HF 612 | x10 | 500 MHz | < 0.7 ns | 10 MΩ, 10 pF | £150 |
Typical Applications
Signal integrity checks on 50 MHz to 100 MHz SPI and QSPI flash interfaces. Gate drive and dead-time measurement on fast switching converters. Matching probes to a 350 MHz bench oscilloscope.
The TT-HX 212 suits engineers who work with 200 MHz to 350 MHz oscilloscopes and want a passive probe at that bandwidth, and who keep a separate x1 probe for low-level work. Browse all TESTEC passive oscilloscope probes.
Order the TESTEC TT-HX 212 from £60 ex VAT. Ask us to check compatibility with your oscilloscope if its input capacitance is outside 8 pF to 25 pF.
Last reviewed: 23 November 2026
Sources (2 TESTEC documents and pages)
- TESTEC 5 mm probes product page (testec.de)
- TESTEC TT-HX 212/312 instruction manual (in the downloads on this page)
| General | |
|---|---|
| Order number | 12110 |
| Probe type | Passive oscilloscope probe |
| Attenuation | x10 (10:1) |
| Bandwidth | 350 MHz |
| Rise time | 1 ns |
| Input | |
| Input impedance | 10 MΩ, 13 pF |
| Maximum input voltage | 400 Vrms CAT II; 1500 V transient overvoltage (derating with frequency) |
| Compensation range | 8 pF to 25 pF |
| Accuracy and noise | |
| DC accuracy | ±2% |
| Connection and power | |
| Cable length | 1.2 m |
| Safety and environment | |
| Safety standard | EN 61010-031 |
| Pollution degree | 2 |
| Operating temperature | 0 °C to 50 °C |
| Humidity | 5% to 95% RH (10 °C to 30 °C); 5% to 50% RH above 40 °C |
| Operating altitude | Up to 3000 m |
| Ordering information | |
| Part number (SKU) | 12110 |
| Manufacturer | TESTEC Elektronik GmbH |
| EAN | 5055383603701 |
| Warranty | 12 months |
| Shipping weight | 0.5 kg |
-
1 ×
TESTEC TT-HX 212 350MHz Passive Oscilloscope Probe
Contact us to confirm the full contents of the box.
Questions and Answers
-
What Bandwidth Is the TT-HX 212?
350 MHz with a 1 ns rise time, at fixed x10 attenuation.
-
What Is the Input Impedance (TT-HX 212)?
10 MΩ in parallel with 13 pF at the probe tip.
-
What Is the Maximum Input Voltage (TT-HX 212)?
400 Vrms Measurement Category II with 1500 V transient overvoltage, derating with frequency.
No questions match your search.