TESTEC Elektronik GmbH
TESTEC TT-LF 312 150MHz Passive Oscilloscope Probe
TESTEC TT-LF 312 150MHz Passive Oscilloscope Probe

Key Features
Overview
The TESTEC TT-LF 312 is a switchable x1/x10 modular passive oscilloscope probe with 150 MHz bandwidth and 2.3 ns rise time at x10, and 15 MHz at x1. Input loading is 10 MΩ, 15.5 pF at x10 and 47 pF at x1, and the compensation trimmer covers oscilloscope inputs from 10 pF to 60 pF. It is a general-purpose probe for oscilloscopes up to about 150 MHz.
Downloads
Brochures
The TESTEC TT-LF 312 is a switchable x1/x10 modular passive oscilloscope probe with 150 MHz bandwidth and 2.3 ns rise time at x10, and 15 MHz at x1. Input loading is 10 MΩ, 15.5 pF at x10 and 47 pF at x1, and the compensation trimmer covers oscilloscope inputs from 10 pF to 60 pF. It is a general-purpose probe for oscilloscopes up to about 150 MHz.
Switchable x1/x10 with a Wide 10 pF to 60 pF Compensation Range
The 10 pF to 60 pF range is wider than that of TESTEC's newer LX series (8 pF to 25 pF), so the TT-LF 312 can be matched to older or less common oscilloscopes with higher input capacitance. The trimmer sits in the probe body; TESTEC's procedure is to apply a 1 kHz square wave and adjust for the best square-wave response.
150 MHz at x10, Full Sensitivity at x1
At x10 the probe gives 150 MHz, 2.3 ns rise time and a 10 MΩ input. At x1 it passes the full signal at up to 15 MHz with a 24 ns rise time; input resistance is then the same as the oscilloscope's. Set the oscilloscope's probe factor to match the switch position.
Voltage Rating
TESTEC's datasheet gives a maximum input of 600 V (DC plus peak AC) CAT I, derating with frequency. In the x1 position the maximum is the same as the oscilloscope's input rating. Use it only with a grounded oscilloscope, and never dismantle the probe while it is connected to a voltage source.
Limitations
- CAT I rating only. For CAT II work choose a current LX, MX or HX series probe such as the TT-LX 312 (400 Vrms CAT II at x10).
- TESTEC no longer lists the LF series on testec.de; the TT-LX 312 is the current equivalent in the 5 mm range.
- Ground-referenced only. Floating measurements need an active differential probe.
TT-LF 312 Compared with Other TESTEC Switchable Probes
| Model | Attenuation | Bandwidth | Compensation range | Voltage rating (x10) | Price |
|---|---|---|---|---|---|
| TT-LF 312 (this product) | x1 / x10 | 15 MHz / 150 MHz | 10 to 60 pF | 600 V CAT I | £25 |
| TT-MF 312 | x1 / x10 | 20 MHz / 250 MHz | 10 to 60 pF | 600 V CAT I | £25 |
| TT-LX 312 | x1 / x10 | 10 MHz / 150 MHz | 8 to 25 pF | 400 Vrms CAT II | £35 |
| TT-MX 312 | x1 / x10 | 10 MHz / 250 MHz | 8 to 25 pF | 400 Vrms CAT II | £40 |
Typical Applications
Spare or replacement probes for older oscilloscopes whose input capacitance sits above 25 pF. General bench fault-finding on low-voltage circuits up to 150 MHz. Teaching labs that need a low-cost switchable probe.
The TT-LF 312 suits engineers maintaining older oscilloscopes or needing a low-cost switchable probe for CAT I circuits. Browse all TESTEC passive oscilloscope probes.
Order the TESTEC TT-LF 312 from £25 ex VAT. If you need a CAT II rating, ask us about the TT-LX 312 instead.
Last reviewed: 23 November 2026
Sources (2 TESTEC documents and pages)
| General | |
|---|---|
| Order number | 10020 |
| Probe type | Passive oscilloscope probe |
| Attenuation | x1 / x10, switchable |
| Bandwidth | x1: 15 MHz; x10: 150 MHz |
| Rise time | x1: 24 ns; x10: 2.3 ns |
| Input | |
| Input impedance | x1: same as oscilloscope, 47 pF; x10: 10 MΩ, 15.5 pF |
| Maximum input voltage | 600 V (DC + peak AC) CAT I, derating with frequency; x1: same as oscilloscope |
| Compensation range | 10 pF to 60 pF |
| Connection and power | |
| Cable length | 1.2 m |
| Ordering information | |
| Part number (SKU) | 10020 |
| Manufacturer | TESTEC Elektronik GmbH |
| EAN | 5055383622764 |
| Warranty | 12 months |
| Shipping weight | 0.5 kg |
-
1 ×
TESTEC TT-LF 312 150MHz Passive Oscilloscope Probe
Contact us to confirm the full contents of the box.
Questions and Answers
-
What Is the Compensation Range of the TT-LF 312?
10 pF to 60 pF, so it suits oscilloscopes with higher input capacitance than TESTEC's LX series covers.
-
What Voltage Rating Does It Have (TT-LF 312)?
600 V (DC plus peak AC) CAT I, derating with frequency. In the x1 position the limit is the oscilloscope's own input rating.
-
Is There a Current Replacement (TT-LF 312)?
TESTEC's current 5 mm range lists the TT-LX 312, a switchable x1/x10, 150 MHz probe rated 400 Vrms CAT II at x10.