Overview
The OMICRON Lab Test Card Set - Shunt-Thru with Rs is a replacement set of 32 green test cards and one calibrator for the Shunt-Thru with Rs setup of the B-TCA test card adapter. Two user-fitted series resistors set the range, from 1 mΩ to 100 Ω at 0 Ω up to 20 mΩ to 4,500 Ω at 950 Ω, with up to 60 V DC at the DUT. The cards fit the 100-pin card-edge connector of the B-TCA, used with a Bode 100 or Bode 500.
Downloads
User Manuals
The OMICRON Lab Test Card Set - Shunt-Thru with Rs (P0011466) is a replacement set of 32 test cards and one calibrator for the Shunt-Thru with Rs setup of the B-TCA test card adapter. Two series resistors (Rs) fitted by the user set the measurement range: from 1 mΩ to 100 Ω with Rs at 0 Ω up to 20 mΩ to 4,500 Ω with Rs at 950 Ω, within the B-TCA's recommended DC to 450 MHz range, on a Bode 500 (10 mHz to 450 MHz) or Bode 100 (1 Hz to 50 MHz) vector network analyser. Higher Rs values also raise the permitted DC voltage at the DUT, from 3.3 V at 0 Ω to the 60 V safety limit at 950 Ω.
Choosing Rs: Shunt-Thru Impedance Range and DC Voltage
With Rs at 0 Ω the setup behaves like plain Shunt-Thru; larger Rs values extend it towards higher impedance. The B-TCA user manual gives these optimal ranges and limits:
| Rs | Optimal impedance range | Max. DC voltage at DUT | Max. power in Rs |
|---|---|---|---|
| 0 Ω | 1 mΩ to 100 Ω | 3.3 V | Not applicable |
| 200 Ω | 5 mΩ to 1,125 Ω | 16.5 V | 0.9 W |
| 450 Ω | 10 mΩ to 2,250 Ω | 33 V | 2 W |
| 499 Ω | 11 mΩ to 2,480 Ω | 36 V | 2.2 W |
| 950 Ω | 20 mΩ to 4,500 Ω | 60 V (safety limit) | ≈ 4 W |
Power in Rs follows P = (V / (Rs + 50 Ω))² × Rs, where V is the voltage at the DUT connection point.
Fitting Rs and Calibrating the Green Calibrator
On each green test card, solder the DUT between the signal pad and the GND pad, then solder two Rs resistors (0603 to 1206) at the marked locations to close the measurement path. The maximum recommended DUT size is about 9 mm x 9 mm. Enter the Rs value in the Impedance/Reflection Measurement Setup or the Calibration dialog of the Bode Analyzer Suite, then run Open, Short with the Short-Delay Time set to 23 ps, and Load (50 Ω ±0.1%).
Soldering and Fitting B-TCA Test Cards
Each test card is a 1.57 mm (nominal) PCB for the 100-pin, 0.8 mm pitch card-edge connector of the B-TCA test card adapter, with the DUT soldered directly to the marked pads. Signal line contact resistance is 2 mΩ typ. with under 1 mΩ repeatability, and the connection is rated for 1,000 mating cycles. Wear gloves when breaking cards off and remove any sharp edges. Clean the contacts, then push the card straight down at its centre with under 2° of tilt.
Specifications
| Order number | P0011466 |
|---|---|
| Measurement setup | Shunt-Thru with Rs (two-port impedance) |
| Test card colour | Green |
| DUT connection | Between signal and GND pads, plus 2 Rs solder locations |
| Rs package sizes | 0603 to 1206 |
| Best-suited impedance range | Depends on Rs: 1 mΩ to 100 Ω (0 Ω) up to 20 mΩ to 4,500 Ω (950 Ω) |
| Max. recommended DUT size | ≈ 9 mm x 9 mm |
| Calibrator references | Open, Short, Load (50 Ω ±0.1%), 6 Rs locations |
| Short-Delay Time | 23 ps |
| Recommended frequency range | DC to 450 MHz (B-TCA) |
| Card thickness | 1.57 mm (nominal) |
| Card-edge connector | 100 pins, 0.8 mm pitch |
| Signal line contact resistance (4 pins) | 2 mΩ typ., < 1 mΩ repeatability typ. |
| Test card ground contact resistance | < 0.5 mΩ typ., < 100 µΩ repeatability typ. |
| Test card mating cycles | 1,000 |
| Max. voltage and current (B-TCA) | 60 V DC, 30 V AC, 1 A |
| Fits | B-TCA test card adapter (P0009653) |
| Compatible analysers | Bode 100, Bode 500 (via the B-TCA) |
Limitations
- These are replacement parts for the B-TCA test card adapter. The adapter and the Bode 100 or Bode 500 analyser are not included.
- B-TCA limits are 60 V DC or 30 V AC, 1 A, and under 350 mJ stored in an inductive DUT. Use only DC sources isolated from mains.
- The B-TCA has no internal bias-T. Above 3.3 V DC, OMICRON Lab suggests an external bias-T between the analyser and the adapter.
- Rs resistors are not on the OMICRON Lab delivery list for this set. For 1 kΩ to 1 MΩ without resistors, choose the Series-Thru set.
B-TCA Test Card Set Comparison
| Feature | B-TCA adapter | Shunt-Thru set | Series-Thru set | One-Port set | Shunt-Thru with Rs set |
|---|---|---|---|---|---|
| Order number | P0009653 | P0011462 | P0011464 | P0011465 | P0011466 |
| Contents | Adapter, 2 N-N and 2 N-BNC cables, 4 calibrators, 32 test cards (8 per setup) | 32 test cards, 1 calibrator | 32 test cards, 1 calibrator | 32 test cards, 1 calibrator | 32 test cards, 1 calibrator |
| Test card colour | All four | Blue | Red | Black | Green |
| Best-suited impedance range | Milliohms to megohms across four setups | 1 mΩ to 100 Ω | 1 kΩ to 1 MΩ | 500 mΩ to 10 kΩ | Set by Rs: 1 mΩ to 100 Ω (0 Ω) up to 20 mΩ to 4,500 Ω (950 Ω) |
| Max. recommended DUT size | Per setup | ≈ 9 mm contact to contact, 10 mm wide | ≈ 9 mm contact to contact | 15 mm contact to contact | ≈ 9 mm x 9 mm |
| Short-Delay Time | Per setup | 23 ps | 0 ps | 0 ps | 23 ps |
| Price | £950 | £175 | £165.02 | £175 | £175 |
Typical Applications
Impedance against frequency, up to 450 MHz, of SMD parts between 20 mΩ and 4,500 Ω with a 950 Ω Rs, beyond the 100 Ω ceiling of plain Shunt-Thru. Measurements with a DC voltage at the DUT above 3.3 V and up to 60 V, set by the Rs value.
This set suits engineers who already run a B-TCA and have used up the eight Shunt-Thru with Rs cards supplied with it.
What's Included
- 32 Shunt-Thru with Rs test cards (4 × 8)
- 1 Shunt-Thru with Rs calibrator
The OMICRON Lab Test Card Set - Shunt-Thru with Rs is £175 ex VAT. If you do not have the adapter yet, the B-TCA ships with eight cards and a calibrator per setup; contact us for a formal quotation.
Frequently Asked Questions
Which Rs Value Should I Fit?
Pick the Rs whose optimal range covers your DUT and whose DC voltage limit covers your bias. For example, 200 Ω gives 5 mΩ to 1,125 Ω with up to 16.5 V DC at the DUT, and 950 Ω gives 20 mΩ to 4,500 Ω up to the 60 V safety limit. Use the same Rs value on the calibrator and the test cards.
Answered by John Legg, Director, Debug Store
Are the Rs Resistors Included?
No. OMICRON Lab lists 32 test cards and one calibrator for this set. You fit two Rs resistors (0603 to 1206) to each test card and six to the calibrator.
Answered by John Legg, Director, Debug Store
Last reviewed: 18 November 2026
Sources (1 manufacturer page, 1 user manual)
| Ordering information | |
|---|---|
| Part number (SKU) | P0011466 |
| Manufacturer | OMICRON-Lab |
| EAN | 5055383604081 |
| Warranty | 12 months |
| Shipping weight | 1.0 kg |
- 1 ×Calibrator for Shunt-Thru with Rs
- 32 ×Shunt-Thru with Rs test card