TESTEC Elektronik GmbH
TESTEC TT-HVP 15 B - 10 kV DC / 40 MHz High Voltage Oscilloscope Probe
TESTEC TT-HVP 15 B - 10 kV DC / 40 MHz High Voltage Oscilloscope Probe

Hauptmerkmale
Übersicht
The TESTEC TT-HVP 15 B is a 1000:1 passive high voltage oscilloscope probe rated 20 kV (DC + AC peak), 10 kV DC and 7 kV AC RMS, with 40 MHz bandwidth, a rise time of 8.8 ns or less and a 100 MΩ, 3.0 pF input. It is designed for Measurement Category 0 laboratory work on high-voltage supplies and pulse circuits, and connects to a standard oscilloscope BNC input with no external power.
The TESTEC TT-HVP 15 B is a 1000:1 passive high voltage oscilloscope probe rated 20 kV (DC + AC peak), 10 kV DC and 7 kV AC RMS, with 40 MHz bandwidth, a rise time of 8.8 ns or less and a 100 MΩ, 3.0 pF input. It is designed for Measurement Category 0 laboratory work on high-voltage supplies and pulse circuits, and connects to a standard oscilloscope BNC input with no external power.
20 kV Peak with 40 MHz Bandwidth
At 1000:1, 20 kV peak at the tip reads as 20 V on the oscilloscope. With 3.0 pF input capacitance and 40 MHz bandwidth, the probe captures pulse edges as well as DC levels. Signal to noise is better than 60 dB at 1 kHz and 50 dB at 1 MHz, and the temperature coefficient is 200 ppm/°C or less. DC accuracy is within 6%.
Duty Limits Above 8 kV DC
TESTEC limits continuous use at the top of the range. Above 8 kV DC, 5 kV AC RMS or 15 kV peak, keep contact with the circuit to 60 seconds, with 10 minutes between measurements. Plan measurements on high-voltage supplies around this, or choose the TT-HVP 2739 for higher headroom.
Compensation, Safety and Environment
The compensation range is 15 pF to 30 pF. TESTEC builds the probe to IEC 61010-031:2015 for Measurement Category 0, pollution degree 2, at altitudes up to 4600 m. Measurements are ground-referenced only. The cable is 2 m (±0.2 m), and the probe measures 75 × 75 × 340 mm and weighs 250 g.
Limitations
- Measurement Category 0 only; not for mains-connected installations.
- The voltage rating derates with frequency; check the curve in TESTEC's manual before measuring fast high-voltage signals.
- Above 8 kV DC, 5 kV AC RMS or 15 kV peak, contact is limited to 60 seconds with 10 minutes between measurements.
- Ground-referenced only; floating measurements are not permitted.
- For up to 39 kV peak choose the TT-HVP 2739; for a multimeter, the TT-HVP 40.
TT-HVP 15 B Compared with Other TESTEC High Voltage Probes
| Model | Ratio | Maximum voltage | Bandwidth | Input | Price |
|---|---|---|---|---|---|
| TT-HV 400 | 100:1 | 4000 V (DC + peak AC) | 300 MHz | 100 MΩ, 5 pF | £60 |
| TT-HVP 08 | 1000:1 | 8 kV DC; 16 kV peak | 40 MHz | 100 MΩ, 3 pF | £275 |
| TT-HVP 15 B (this product) | 1000:1 | 20 kV (DC + AC peak) | 40 MHz | 100 MΩ, 3 pF | £390 |
| TT-HVP 2739 | 1000:1 | 39 kV (DC + AC peak) | 50 MHz | 900 MΩ, 2 pF | £1,025 |
| TT-HVP 40 | 1000:1 (into 10 MΩ) | 40 kV DC | 300 Hz (-3 dB) | 1000 MΩ (meter probe) | £145 |
Typical Applications
Characterising output and ripple of 10 kV to 20 kV laboratory supplies. Capturing pulses from capacitor discharge and high-voltage pulse circuits. Verifying voltage waveforms in dielectric and insulation test fixtures.
The TT-HVP 15 B suits lab engineers who need oscilloscope views of signals into the tens of kilovolts. See all high voltage probes.
Order the TESTEC TT-HVP 15 B from £390 ex VAT. If you are choosing between this and the TT-HVP 2739, tell us your peak voltage and duty and we will advise.
Last reviewed: 23 November 2026
Sources (3 TESTEC documents and pages)
- TESTEC TT-HVP 15 B datasheet (PDF)
- TESTEC high voltage probes product page (testec.de)
- TESTEC TT-HVP 15 B instruction manual (in the downloads on this page)
| General | |
|---|---|
| Order number | 15010 |
| Probe type | High voltage oscilloscope probe |
| Attenuation | 1000:1 |
| Bandwidth | 40 MHz |
| Rise time | ≤ 8.8 ns |
| Input | |
| Input impedance | 100 MΩ, 3.0 pF ±0.5 pF |
| Maximum input voltage | 20 kV DC + AC peak (CAT 0); 10 kV DC; 7 kV AC RMS |
| Compensation range | 15 pF to 30 pF |
| Accuracy and noise | |
| DC accuracy | ≤ 6% |
| Temperature coefficient | ≤ 200 ppm/°C |
| Signal-to-noise ratio | > 60 dB at 1 kHz; > 50 dB at 1 MHz |
| Connection and power | |
| Cable length | 2 m ±0.2 m |
| Safety and environment | |
| Safety standard | IEC 61010-031:2015, CAT 0, pollution degree 2 |
| Operating altitude | Up to 4600 m |
| Mechanical | |
| Dimensions | 75 × 75 × 340 mm |
| Weight | 250 g |
| Bestellinformationen | |
| Artikelnummer (SKU) | 15010 |
| Hersteller | TESTEC Elektronik GmbH |
| EAN | 5055383622726 |
| Garantie | 12 months |
| Versandgewicht | 1.5 kg |
-
1 ×
TESTEC TT-HVP 15 B - 10 kV DC / 40 MHz High Voltage Oscilloscope Probe
Kontaktieren Sie uns, um den vollständigen Lieferumfang zu bestätigen.
Fragen und Antworten
-
What Is the Maximum Voltage of the TT-HVP 15 B?
20 kV DC + AC peak, 10 kV DC and 7 kV AC RMS, in Measurement Category 0.
-
Can It Stay Connected Continuously (TT-HVP 15 B)?
Not at the top of its range. Above 8 kV DC, 5 kV AC RMS or 15 kV peak, TESTEC limits contact to 60 seconds with 10 minutes between measurements.
-
What Compensation Range Does It Cover (TT-HVP 15 B)?
15 pF to 30 pF oscilloscope input capacitance, on a 1 MΩ input.