Overview
The OMICRON Lab Test Card Set - Shunt-Thru is a replacement set of 32 blue Shunt-Thru test cards and one Shunt-Thru calibrator (Open, Short, Load) for the B-TCA test card adapter. Shunt-Thru is rated best of the four B-TCA setups for low impedance, 1 mΩ to 100 Ω, within the B-TCA's DC to 450 MHz range (up to 50 MHz on a Bode 100), with DUTs up to about 9 mm contact to contact. The cards fit the 100-pin card-edge connector of the B-TCA, used with a Bode 100 or Bode 500.
Download
User Manuals
- Details
OMICRON Lab B-TCA Test Card Adapter
The OMICRON Lab Test Card Set - Shunt-Thru (P0011462) is a replacement set of 32 Shunt-Thru test cards and one Shunt-Thru calibrator for the B-TCA test card adapter. OMICRON Lab rates Shunt-Thru as the best of the four B-TCA setups for low impedance, with an optimal DUT impedance magnitude of 1 mΩ to 100 Ω, within the B-TCA's recommended DC to 450 MHz range, on a Bode 500 (10 mHz to 450 MHz) or Bode 100 (1 Hz to 50 MHz) vector network analyser. Order it to restock the blue Shunt-Thru cards when you characterise low-impedance passive SMD components.
Shunt-Thru Impedance Measurement From 1 mΩ to 100 Ω
On a Shunt-Thru test card the DUT is soldered between the signal pad and the GND pad, so it shunts the signal line to ground. This two-port arrangement gives the best low-impedance sensitivity of the four setups. Above 100 Ω the 50 Ω parallel path degrades sensitivity, so for high-impedance parts move to the Series-Thru set. The maximum recommended DUT size is about 9 mm from one contact to the other, with a maximum width of 10 mm.
Shunt-Thru Calibrator: Open, Short and Load
The blue Shunt-Thru calibrator carries Open (Thru), Short and Load references, with the Load at 50 Ω ±0.1%. In the Bode Analyzer Suite, run Open first, then Short with the Short-Delay Time set to 23 ps, then Load. A Thru-only calibration using the Open element is possible, but OMICRON Lab notes that it does not remove impedance mismatch or the low-frequency ground-loop error, so a full Open, Short and Load calibration is the one to use for low-impedance work.
Soldering and Fitting B-TCA Test Cards
Each test card is a 1.57 mm (nominal) PCB for the 100-pin, 0.8 mm pitch card-edge connector of the B-TCA test card adapter, with the DUT soldered directly to the marked pads. Signal line contact resistance is 2 mΩ typ. with under 1 mΩ repeatability, and the connection is rated for 1,000 mating cycles. Wear gloves when breaking cards off and remove any sharp edges. Clean the contacts, then push the card straight down at its centre with under 2° of tilt.
Specifications
| Order number | P0011462 |
|---|---|
| Measurement setup | Shunt-Thru (two-port impedance) |
| Test card colour | Blue |
| DUT connection | Between signal and GND pads |
| Best-suited impedance range | 1 mΩ to 100 Ω (impedance magnitude) |
| Max. recommended DUT size | ≈ 9 mm contact to contact, max. width 10 mm |
| Calibrator references | Open (Thru), Short, Load (50 Ω ±0.1%) |
| Short-Delay Time | 23 ps |
| Recommended frequency range | DC to 450 MHz (B-TCA) |
| Card thickness | 1.57 mm (nominal) |
| Card-edge connector | 100 pins, 0.8 mm pitch |
| Signal line contact resistance (4 pins) | 2 mΩ typ., < 1 mΩ repeatability typ. |
| Test card ground contact resistance | < 0.5 mΩ typ., < 100 µΩ repeatability typ. |
| Test card mating cycles | 1,000 |
| Max. voltage and current (B-TCA) | 60 V DC, 30 V AC, 1 A |
| Fits | B-TCA test card adapter (P0009653) |
| Compatible analysers | Bode 100, Bode 500 (via the B-TCA) |
Limitations
- These are replacement parts for the B-TCA test card adapter. The adapter and the Bode 100 or Bode 500 analyser are not included.
- B-TCA limits are 60 V DC or 30 V AC, 1 A, and under 350 mJ stored in an inductive DUT. Use only DC sources isolated from mains.
- The B-TCA has no internal bias-T. Above 3.3 V DC, OMICRON Lab suggests an external bias-T between the analyser and the adapter.
- Sensitivity degrades above 100 Ω. For 1 kΩ to 1 MΩ choose the Series-Thru set; to push a shunt measurement up to 4,500 Ω, use the Shunt-Thru with Rs set with a 950 Ω Rs.
B-TCA Test Card Set Comparison
| Feature | B-TCA adapter | Shunt-Thru set | Series-Thru set | One-Port set | Shunt-Thru with Rs set |
|---|---|---|---|---|---|
| Order number | P0009653 | P0011462 | P0011464 | P0011465 | P0011466 |
| Contents | Adapter, 2 N-N and 2 N-BNC cables, 4 calibrators, 32 test cards (8 per setup) | 32 test cards, 1 calibrator | 32 test cards, 1 calibrator | 32 test cards, 1 calibrator | 32 test cards, 1 calibrator |
| Test card colour | All four | Blue | Red | Black | Green |
| Best-suited impedance range | Milliohms to megohms across four setups | 1 mΩ to 100 Ω | 1 kΩ to 1 MΩ | 500 mΩ to 10 kΩ | Set by Rs: 1 mΩ to 100 Ω (0 Ω) up to 20 mΩ to 4,500 Ω (950 Ω) |
| Max. recommended DUT size | Per setup | ≈ 9 mm contact to contact, 10 mm wide | ≈ 9 mm contact to contact | 15 mm contact to contact | ≈ 9 mm x 9 mm |
| Short-Delay Time | Per setup | 23 ps | 0 ps | 0 ps | 23 ps |
| Price | £950 | £175 | £165.02 | £175 | £175 |
Typical Applications
Impedance against frequency, up to 450 MHz, of SMD parts whose impedance magnitude sits between 1 mΩ and 100 Ω. Low-impedance measurements where the One-Port setup's 500 mΩ lower limit is too high.
This set suits engineers who already run a B-TCA and have used up the eight Shunt-Thru cards supplied with it.
What's Included
- 32 Shunt-Thru test cards (4 × 8)
- 1 Shunt-Thru calibrator
The OMICRON Lab Test Card Set - Shunt-Thru is £175 ex VAT. If you do not have the adapter yet, the B-TCA ships with eight cards and a calibrator per setup; contact us for a formal quotation.
Frequently Asked Questions
Why Choose Shunt-Thru Over One-Port for Low Impedance?
Because its best-suited range starts at 1 mΩ, against 500 mΩ for One-Port. OMICRON Lab rates Shunt-Thru best and One-Port only OK for low impedance, as the One-Port range is limited by contact resistance and the analyser's signal-to-noise ratio.
Answered by John Legg, Director, Debug Store
Is This the Same as the Shunt-Thru with Rs Set?
No. Shunt-Thru with Rs cards are green and carry extra solder locations for two series resistors (Rs, 0603 to 1206), which move the measurement range upwards. With Rs at 0 Ω that setup behaves like Shunt-Thru, but the blue Shunt-Thru cards here need no resistors.
Answered by John Legg, Director, Debug Store
Last reviewed: 18 November 2026
Sources (1 manufacturer page, 1 user manual)
| Ordering information | |
|---|---|
| Part number (SKU) | P0011462 |
| Manufacturer | OMICRON-Lab |
| EAN | 5055383603985 |
| Warranty | 12 months |
| Shipping weight | 1.0 kg |
- 1 ×Calibrator for Shunt-Thru
- 32 ×Shunt-Thru test card