Caratteristiche principali
Overview
The OMICRON Lab DSH 100 is a dielectric sample holder for solid materials such as polymers, epoxy resins, insulation papers, glasses and thin films, with a guarded parallel plate electrode system. Usable from 5 µHz to 5 MHz at up to 200 V peak, it takes samples 0.1 mm to 20 mm thick, operates from -55 °C to +125 °C and supports ASTM D150 and IEC 62631-2-1. Connect it to a Bode 100 or BNC LCR meter with the Triaxial to BNC Adapter Cable.
Download
User Manuals
- Details
OMICRON Lab DSH 100 Dielectric Sample Holder
Perché gli ingegneri scelgono OMICRON-Lab DSH 100 Dielectric Sample Holder
Precision Material Characterisation
Standards Compliance Assurance
Versatile Integration Capability
The OMICRON Lab DSH 100 is a dielectric sample holder for solid insulating materials, with a guarded parallel plate electrode system usable from 5 µHz to 5 MHz at up to 200 V peak. It takes samples 0.1 mm to 20 mm thick and 50 x 50 mm to 70 x 70 mm, and operates from -55 °C to +125 °C. Materials and insulation engineers use it with a Bode 100, impedance analyser or LCR meter to measure polymers, epoxy resins, insulation papers, glasses and thin films to ASTM D150 and IEC 62631-2-1.
Guarded Parallel Plate Electrodes for Solid Dielectrics
The top (IN) electrode is a 70 mm diameter PCB disc. The bottom (MEAS) electrode is 49 mm in diameter and surrounded by a 9.5 mm guard ring with a 1 mm gap. The electrodes are exchangeable and low cost, which OMICRON Lab notes makes it practical to analyse curing processes, for example in epoxy systems. Two 0.3 mm and two 0.8 mm spacers are supplied for air-reference measurements.
Controlled Clamping and Temperature Testing
A tensioner controls the pressure force on the sample, with 50 N and 100 N spring systems supplied. The housing has mounting provision for a heating pad, a pressure sensor and a temperature sensor (not supplied), and the holder operates from -55 °C to +125 °C at up to 95% relative humidity, non-condensing. OMICRON Lab warns that the metal surface must not be touchable while operating above 65 °C.
Connecting the Dielectric Sample Holder to an Analyser
The DSH 100 connects through a triaxial LEMO interface (female plug PSA.1S.650.CLL). MEAS goes to the analyser input and IN to the analyser output. The Triaxial to BNC Adapter Cable (B1023000) connects it to the Bode 100 or any LCR meter with BNC connections. Use only a signal source that provides a galvanically isolated (floating) voltage, stay within 200 V peak and 50 mA peak, and connect the grounding terminal to laboratory earth. A safety interlock stops the holder opening while a cable is connected.
Specifications
| Frequency range | 5 µHz to 5 MHz |
|---|---|
| Max. voltage (AC/DC) | ≤ 200 V peak |
| Max. current (AC/DC) | ≤ 50 mA peak |
| Connector | Triaxial, LEMO female plug PSA.1S.650.CLL |
| Sample thickness | 0.1 mm to 20 mm |
| Sample size | 50 x 50 mm to 70 x 70 mm |
| Top electrode | Ø 70 mm |
| Bottom electrode (with guard ring) | Ø 49 mm |
| Guard ring width / gap | 9.5 mm / 1 mm |
| Spacers (air reference) | 2 x 0.8 mm, 2 x 0.3 mm |
| Spring force (clamping system) | 50 N and 100 N |
| Operating temperature | -55 °C to +125 °C |
| Relative humidity | ≤ 95%, non-condensing |
| Max. altitude | 2,000 m |
| Dimensions (W x H x D) | 165 x 108 x 118 mm |
| Weight | 2.5 kg |
| Shielding | Full metal shielding |
| Safety | Safety interlock |
| Standards | ASTM D150; IEC 62631-2-1 (2018); IEC 62631-3-1 (2016) |
| Order number | P0005783 |
Limitations
- The analyser and the B1023000 adapter cable are not included.
- The analyser sets the usable band. The Bode 100 covers 1 Hz to 50 MHz, so with it the DSH 100 runs from 1 Hz to 5 MHz. Reaching 5 µHz needs an analyser such as OMICRON Lab's SPECTANO 100 (5 µHz to 5 kHz), discontinued on 31 December 2022.
- OMICRON Lab does not name the Bode 500 for the DSH 100, so ask us before planning a Bode 500 setup.
- OMICRON Lab designed the DSH 100 for solid materials.
- Replace cables and electrodes after 500 mating cycles.
Bode 100 Test Fixture Comparison
| Feature | DSH 100 | B-WIC | B-SMC |
|---|---|---|---|
| Device under test | Solid dielectric samples, 0.1 mm to 20 mm thick | Wired (through-hole) components | Surface-mount components |
| Frequency range | 5 µHz to 5 MHz | 1 Hz to 50 MHz (usable) | 1 Hz to 50 MHz (usable) |
| Analysers named by OMICRON Lab | Bode 100, impedance analysers, LCR meters | Bode 100 | Bode 100 |
| Price | £5,215 | £650 | £650 |
Typical Applications
Dielectric response of polymers and thin films across frequency, with the supplied spacers for the air-reference measurement. Temperature-dependent measurement of insulation papers and glasses between -55 °C and +125 °C, with a heating pad and temperature sensor fitted, or with the holder in a climate chamber for sub-ambient tests.
The DSH 100 suits insulation and materials engineers who already have a Bode 100 or LCR meter.
What's Included
- 1 x DSH 100 Dielectric Sample Holder
- 2 x PCB disk electrodes, one electrode with guard ring
- 2 x Tension control systems, with 50 N and 100 N spring
- 2 x Metal top plates, 10 mm and 5 mm thick
- 2 x Spacers, 0.3 mm thick
- 2 x Spacers, 0.8 mm thick
- 1 x Carrying case
- 1 x User manual, including safety instructions
The OMICRON Lab DSH 100 is £5,215 ex VAT, and the Triaxial to BNC Adapter Cable is £125. Contact us for a formal quotation covering the holder, cable and a Bode 100.
Frequently Asked Questions
Does the DSH 100 Work with the Bode 100?
Yes. OMICRON Lab names the Bode 100 as a compatible analyser and supplies the B1023000 Triaxial to BNC Adapter Cable to connect it. With the Bode 100, the measurement range is 1 Hz to 5 MHz.
Answered by John Legg, Director, Debug Store
What Does the Analyser Source Need to Provide?
A galvanically isolated (floating) voltage, isolated from mains, at no more than 200 V peak and 50 mA peak. Connect the DSH 100 grounding terminal to laboratory earth before measuring.
Answered by John Legg, Director, Debug Store
Last reviewed: 18 November 2026
| Ordering information | |
|---|---|
| Part number (SKU) | P0005783 |
| Manufacturer | OMICRON-Lab |
| EAN | 5055383604449 |
| Warranty | 12 months |
| Shipping weight | 4.0 kg |
- 1 ×DSH 100 Dielectric Sample Holder
- 2 ×PCB disk electrodes, one with guard ring
- 2 ×Tension control systems, 50 N and 100 N spring
- 2 ×Metal top plates, 10 mm and 5 mm thick
- 2 ×Spacers, 0.3 mm thick
- 2 ×Spacers, 0.8 mm thick
- 1 ×Carrying case
- 1 ×User manual, including safety instructions
Domande e risposte
-
How Does the Guard Ring Electrode Configuration Improve Measurement Accuracy?
The integrated guard ring reduces fringing field effects, and stray capacitance is corrected with an air-reference measurement using the supplied spacers. This matters most for low-loss materials and precise permittivity measurements.
-
What Temperature Range Is Supported for Material Testing Applications?
The DSH 100 operates from -55°C to +125°C, so you can characterise materials across a wide temperature range, from low to elevated temperatures.
-
What Frequency and Voltage Range Does the DSH 100 Cover?
The DSH 100 covers 5 µHz to 5 MHz and has a maximum operating voltage of 200 V peak.
-
What Safety Features Are Incorporated in the DSH 100 Design?
The DSH 100 includes:
- a mandatory safety interlock that prevents operation with the housing open
- full metal shielding
- construction designed for operator safety during high-voltage measurements
-
Can the DSH 100 Monitor Epoxy Curing Processes in Real-time?
Yes. The exchangeable electrodes and wide frequency range let the DSH 100 track changes in dielectric properties as epoxy cures from liquid to solid.
-
Which International Standards Does the DSH 100 Comply with for Dielectric Testing?
The DSH 100 supports measurements in accordance with ASTM D150, IEC 62631-2-1 (2018) and IEC 62631-3-1 (2016).
-
What Types of Materials Can the DSH 100 Analyse Effectively?
The DSH 100 is designed for solid dielectric materials used in electronic and electrical applications, including:
- nano-composites
- polymers
- epoxy resins
- insulation papers
- glasses
- thin films
-
What Frequency Range Does the DSH 100 Support for Dielectric Measurements?
The DSH 100 covers 5 µHz to 5 MHz, so you can analyse different polarisation mechanisms and dielectric effects across many frequency decades.
-
Can the DSH 100 Be Used with Different Impedance Analysers Besides the SPECTANO 100?
Yes. Besides the SPECTANO 100, the DSH 100 works with other impedance analysers, LCR meters and the Bode 100 via its LEMO triaxial connectors (an optional triaxial-to-BNC adapter cable is available for the Bode 100 and BNC-equipped LCR meters). The analyser's source must be galvanically isolated.
-
What Sample Sizes and Thicknesses Can the DSH 100 Accommodate?
The DSH 100 takes samples from 50mm x 50mm to 70mm x 70mm, with thickness from 0.1mm to 20mm.
No questions match your search.