TESTEC Elektronik GmbH
Sondes d'oscilloscope passives TESTEC série TT-MF-300 250 MHz
Sondes d'oscilloscope passives TESTEC série TT-MF-300 250 MHz

Caractéristiques principales
Présentation
The TESTEC TT-MF 312 is a switchable x1/x10 modular passive oscilloscope probe with 250 MHz bandwidth and 1.4 ns rise time at x10, and 20 MHz at x1. Input loading is 10 MΩ, 15 pF at x10, compensation covers 10 pF to 60 pF, and a separate high-frequency adjustment in the BNC box sets the pulse response. It suits 100 MHz to 200 MHz oscilloscopes, including older models with higher input capacitance.
Téléchargements
Brochures
The TESTEC TT-MF 312 is a switchable x1/x10 modular passive oscilloscope probe with 250 MHz bandwidth and 1.4 ns rise time at x10, and 20 MHz at x1. Input loading is 10 MΩ, 15 pF at x10, compensation covers 10 pF to 60 pF, and a separate high-frequency adjustment in the BNC box sets the pulse response. It suits 100 MHz to 200 MHz oscilloscopes, including older models with higher input capacitance.
250 MHz at x10 with Separate LF and HF Adjustment
The MF series adds a high-frequency adjustment to the usual low-frequency trimmer. TESTEC's procedure is to set the trimmer in the probe body against a 1 kHz square wave, then set the potentiometer in the BNC connector box against a 1 MHz square wave for the best edge shape. That second step is what lets the probe hold a clean 1.4 ns edge on different oscilloscopes.
Wide 10 pF to 60 pF Compensation Range
The compensation range reaches 60 pF, wider than the 8 pF to 25 pF of TESTEC's newer MX series, so the probe can be matched to older oscilloscopes as well as current ones. At x1 the probe passes the full signal at up to 20 MHz with 47 pF loading; input resistance is then the same as the oscilloscope's.
Voltage Rating
TESTEC's datasheet gives 600 V (DC plus peak AC) CAT I, derating with frequency; at x1 the limit is the oscilloscope's input rating. Use it only with a grounded oscilloscope.
Specifications
| Order number | 11020 |
|---|---|
| Attenuation | x1 / x10, switchable |
| Bandwidth | x1: 20 MHz; x10: 250 MHz |
| Rise time (x10) | 1.4 ns |
| Input impedance | x1: same as oscilloscope, 47 pF; x10: 10 MΩ, 15 pF |
| Compensation range | 10 pF to 60 pF |
| Adjustment | LF trimmer in probe body (1 kHz); HF potentiometer in BNC box (1 MHz) |
| Maximum input voltage | 600 V (DC + peak AC) CAT I, derating with frequency; x1: same as oscilloscope |
| Cable length | 1.2 m |
Limitations
- CAT I rating only. For CAT II work choose the TT-MX 312 (400 Vrms CAT II at x10).
- TESTEC no longer lists the MF series on testec.de; the TT-MX 312 is the current 250 MHz switchable equivalent.
- Ground-referenced only. Floating measurements need an active differential probe.
TT-MF 312 Compared with Other TESTEC Switchable Probes
| Model | Attenuation | Bandwidth | Compensation range | Voltage rating (x10) | Price |
|---|---|---|---|---|---|
| TT-LF 312 | x1 / x10 | 15 MHz / 150 MHz | 10 to 60 pF | 600 V CAT I | £25 |
| TT-MF 312 (this product) | x1 / x10 | 20 MHz / 250 MHz | 10 to 60 pF | 600 V CAT I | £25 |
| TT-MX 312 | x1 / x10 | 10 MHz / 250 MHz | 8 to 25 pF | 400 Vrms CAT II | £40 |
| TT-HX 312 | x1 / x10 | 10 MHz / 350 MHz | 8 to 25 pF | 400 Vrms CAT II | £60 |
Typical Applications
Replacement probes for older 100 MHz to 200 MHz oscilloscopes with input capacitance above 25 pF. Edge and timing checks on low-voltage digital circuits where the HF trim helps match the probe to the instrument. Low-cost probes for teaching labs.
The TT-MF 312 suits engineers who want a 250 MHz switchable probe that can be trimmed to a wide range of oscilloscopes. Browse all TESTEC passive oscilloscope probes.
Order the TESTEC TT-MF 312 from £25 ex VAT. Need a CAT II rating? Ask us about the TT-MX 312.
Last reviewed: 23 November 2026
Sources (2 TESTEC documents and pages)
| Informations de commande | |
|---|---|
| Référence (SKU) | 11020 |
| Fabricant | TESTEC Elektronik GmbH |
| EAN | 5055383622771 |
| Garantie | 12 mois |
| Poids d'expédition | 0.5 kg |
-
1 ×
Sondes d'oscilloscope passives TESTEC série TT-MF-300 250 MHz
Contactez-nous pour confirmer le contenu complet du colis.
Questions et réponses
-
What Bandwidth Does the TT-MF 312 Have?
250 MHz at x10 with a 1.4 ns rise time, and 20 MHz at x1.
-
Why Does It Have an HF Adjustment (TT-MF 312)?
The potentiometer in the BNC box is set against a 1 MHz square wave to optimise the pulse response, after the LF trimmer has been set at 1 kHz.
-
What Is the Voltage Rating (TT-MF 312)?
600 V (DC plus peak AC) CAT I, derating with frequency. At x1 the limit is the oscilloscope's own input rating.