{"product_id":"omicron-lab-b-smc-impedance-adapter-uk","title":"OMICRON Lab B-SMC - Impedance Test Fixture for SM Components","description":"\u003cp\u003eThe OMICRON Lab B-SMC is an impedance test fixture for surface-mount components that turns a Bode 100 into an LCR meter or impedance analyser, usable from 1 Hz to 50 MHz over a recommended impedance range of 20 mΩ to 600 kΩ. A gold-plated needle electrode, set by a knurled nut, clamps SMD parts up to 9.5 mm long and 7.5 mm wide against a gold-plated pad, with a contact resistance of around 0.5 mΩ. Hardware and power electronics engineers use it for RLC-Q measurement of chip resistors, capacitors and inductors across frequency.\u003c\/p\u003e\n\n\u003ch2\u003eSMD Impedance Measurement From 1 Hz to 50 MHz\u003c\/h2\u003e\n\u003cp\u003eOMICRON Lab specifies a usable frequency range of 1 Hz to 50 MHz and recommends the fixture for impedances from 20 mΩ to 600 kΩ. Inside is a resistive measurement bridge with a voltage divider and a current shunt: CH1 receives a signal proportional to the DUT voltage and CH2 a signal proportional to the DUT current. The Bode Analyzer Suite runs it in the Impedance Adapter measurement mode.\u003c\/p\u003e\n\n\u003ch2\u003eNeedle Electrode for Surface-Mount Parts up to 9.5 x 7.5 mm\u003c\/h2\u003e\n\u003cp\u003eThe DUT sits on the electrode pad under the centre needle electrode, and the knurled nut sets the needle-to-pad distance to match the part. The largest DUT the B-SMC takes is 9.5 mm long and 7.5 mm wide. Gold plating brings contact resistance down to around 0.5 mΩ, so clean the needle and pad before Short calibration and low-value measurements.\u003c\/p\u003e\n\n\u003ch2\u003eImpedance Test Fixture Setup and Calibration with the Bode 100\u003c\/h2\u003e\n\u003cp\u003eThree BNC sockets connect the fixture to the \u003ca href=\"\/it\/products\/omicron-lab-bode-100-vector-network-analyser-uk\"\u003eBode 100\u003c\/a\u003e: SOURCE from the analyser output, CH1 and CH2 to the matching inputs. Open, Short and Load calibration is required before measuring. For Open, set the needle-to-pad gap to the DUT size so stray capacitance is compensated. For Short, loosen the knurled nut until the needle touches the electrode. For Load, place a 100 Ω resistor between needle and electrode; six 0603 load resistors (100 Ω ±0.1%, 15 ppm\/°C) are supplied for this. OMICRON Lab characterised typical relative accuracy against an Agilent E4980A precision LCR meter at 10 Hz receiver bandwidth, and publishes the chart in the B-WIC and B-SMC user manual.\u003c\/p\u003e\n\n\u003ch2\u003eSpecifications\u003c\/h2\u003e\n\u003ctable\u003e\n  \u003ctr\u003e\n\u003cth\u003eUsable frequency range\u003c\/th\u003e\n\u003ctd\u003e1 Hz to 50 MHz\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eRecommended impedance range\u003c\/th\u003e\n\u003ctd\u003e20 mΩ to 600 kΩ\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eMeasurement principle\u003c\/th\u003e\n\u003ctd\u003eResistive bridge (voltage divider and current shunt), CH1 voltage, CH2 current\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eConnectors\u003c\/th\u003e\n\u003ctd\u003eBNC sockets: SOURCE, CH1, CH2 (outer conductor connected to housing)\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eDUT contacts\u003c\/th\u003e\n\u003ctd\u003eGold-plated needle electrode and pad with spring mechanics\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eContact resistance\u003c\/th\u003e\n\u003ctd\u003eApprox. 0.5 mΩ\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eMax. DUT size\u003c\/th\u003e\n\u003ctd\u003e9.5 mm length, 7.5 mm width\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eLoad calibration resistors\u003c\/th\u003e\n\u003ctd\u003e0603, 100 Ω ±0.1%, 15 ppm\/°C (6 pcs. supplied)\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eDimensions\u003c\/th\u003e\n\u003ctd\u003e100.5 x 68.2 x 55.5 mm (including connectors)\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eWeight\u003c\/th\u003e\n\u003ctd\u003e0.13 kg\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eOperating temperature\u003c\/th\u003e\n\u003ctd\u003e+5 °C to +40 °C\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eStorage temperature\u003c\/th\u003e\n\u003ctd\u003e-35 °C to +60 °C\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eRelative humidity\u003c\/th\u003e\n\u003ctd\u003e20% to 80%, non-condensing\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eSafety\u003c\/th\u003e\n\u003ctd\u003eSELV device (IEC 60950), protection class III\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eCompliance\u003c\/th\u003e\n\u003ctd\u003eCE, RoHS\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/table\u003e\n\n\u003ch2\u003eLimitations\u003c\/h2\u003e\n\u003cul\u003e\n  \u003cli\u003eThe analyser is not included. OMICRON Lab states the B-SMC is to be used only in conjunction with the Bode 100. Its B-SMC documentation does not cover the Bode 500, so ask us before ordering for a Bode 500 setup.\u003c\/li\u003e\n  \u003cli\u003eDo not apply external voltages or currents. No DC bias rating is published.\u003c\/li\u003e\n  \u003cli\u003eOMICRON Lab publishes a typical relative accuracy chart, not a guaranteed accuracy figure, and states the accuracy rating can change without notice.\u003c\/li\u003e\n  \u003cli\u003eFor leaded parts, choose the \u003ca href=\"\/it\/products\/omicron-lab-b-wic-impedance-adapter-uk\"\u003eB-WIC\u003c\/a\u003e, which takes wires up to 1.8 mm in diameter.\u003c\/li\u003e\n\u003c\/ul\u003e\n\n\u003ch2\u003eImpedance Test Fixture Comparison\u003c\/h2\u003e\n\u003cdiv style=\"overflow-x:auto\"\u003e\u003ctable\u003e\n  \u003ctr\u003e\n\u003cth scope=\"col\"\u003eFeature\u003c\/th\u003e\n\u003cth\u003eB-SMC\u003c\/th\u003e\n\u003cth\u003e\u003ca href=\"\/it\/products\/omicron-lab-b-wic-impedance-adapter-uk\"\u003eB-WIC\u003c\/a\u003e\u003c\/th\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eComponent type\u003c\/th\u003e\n\u003ctd\u003eSurface-mount\u003c\/td\u003e\n\u003ctd\u003eWired (through-hole)\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eDUT limits\u003c\/th\u003e\n\u003ctd\u003eUp to 9.5 mm long, 7.5 mm wide\u003c\/td\u003e\n\u003ctd\u003eWire diameter up to 1.8 mm, spacing min. 3 mm\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eUsable frequency range\u003c\/th\u003e\n\u003ctd\u003e1 Hz to 50 MHz\u003c\/td\u003e\n\u003ctd\u003e1 Hz to 50 MHz\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eRecommended impedance range\u003c\/th\u003e\n\u003ctd\u003e20 mΩ to 600 kΩ\u003c\/td\u003e\n\u003ctd\u003e20 mΩ to 600 kΩ\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eCalibration items supplied\u003c\/th\u003e\n\u003ctd\u003e6 x 0603 load resistors, 100 Ω ±0.1%\u003c\/td\u003e\n\u003ctd\u003eB-CAL calibration board (1 pc.)\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003eWeight\u003c\/th\u003e\n\u003ctd\u003e0.13 kg\u003c\/td\u003e\n\u003ctd\u003e0.16 kg\u003c\/td\u003e\n\u003c\/tr\u003e\n  \u003ctr\u003e\n\u003cth\u003ePrice\u003c\/th\u003e\n\u003ctd\u003e\u003cspan data-ds-price=\"omicron-lab-b-smc-impedance-adapter-uk\"\u003e£650\u003c\/span\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cspan data-ds-price=\"omicron-lab-b-wic-impedance-adapter-uk\"\u003e£650\u003c\/span\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/table\u003e\u003c\/div\u003e\n\n\u003ch2\u003eTypical Applications\u003c\/h2\u003e\n\u003cp\u003eCapacitance and ESR of MLCCs and other chip capacitors, using the 20 mΩ low end of the recommended range. Inductance and Q of SMD power inductors swept up to 50 MHz. Impedance against frequency of chip resistors and ferrite beads, to see where parasitic inductance or capacitance takes over.\u003c\/p\u003e\n\u003cp\u003eThe B-SMC suits engineers who already own a Bode 100 and need repeatable impedance measurements on surface-mount passives without building a test board.\u003c\/p\u003e\n\n\u003ch2\u003eWhat's Included\u003c\/h2\u003e\n\u003cul\u003e\n  \u003cli\u003eB-SMC impedance test fixture\u003c\/li\u003e\n  \u003cli\u003eLoad calibration resistors, 100 Ω ±0.1%, size 0603 (6 pcs.)\u003c\/li\u003e\n\u003c\/ul\u003e\n\n\u003cp\u003eThe OMICRON Lab B-SMC is \u003cspan data-ds-price=\"omicron-lab-b-smc-impedance-adapter-uk\"\u003e£650\u003c\/span\u003e ex VAT. If you also need the analyser, contact us for a formal quotation covering the Bode 100 and fixture. See all \u003ca href=\"\/it\/collections\/vector-network-analyser-accessories\"\u003evector network analyser accessories\u003c\/a\u003e in our range.\u003c\/p\u003e\n\n\u003ch2\u003eFrequently Asked Questions\u003c\/h2\u003e\n\u003ch3\u003eDoes the B-SMC Work with the Bode 500?\u003c\/h3\u003e\n\u003cp\u003eOMICRON Lab's B-SMC documentation names only the Bode 100, and the user manual says to use the fixture only in conjunction with the Bode 100. The Bode 500 also has N-type connectors where the B-SMC uses BNC. Contact us before ordering if your analyser is a Bode 500.\u003c\/p\u003e\n\u003cp\u003e\u003csmall\u003eAnswered by John Legg, Director, Debug Store\u003c\/small\u003e\u003c\/p\u003e\n\u003ch3\u003eDoes the B-SMC Use the B-CAL-3 Calibration Board?\u003c\/h3\u003e\n\u003cp\u003eNo. The B-CAL-3 is the calibration element for the B-WIC. The B-SMC calibrates Short by closing the needle onto the electrode and Load with a 100 Ω resistor between needle and electrode. Six 0603 resistors (100 Ω ±0.1%) ship with the fixture, and OMICRON Lab recommends a precision thin-film 100 Ω resistor with low temperature coefficient for best accuracy.\u003c\/p\u003e\n\u003cp\u003e\u003csmall\u003eAnswered by John Legg, Director, Debug Store\u003c\/small\u003e\u003c\/p\u003e\n\u003ch3\u003eWhat Is the Largest SMD Part the B-SMC Can Hold?\u003c\/h3\u003e\n\u003cp\u003e9.5 mm long and 7.5 mm wide. The needle electrode sits in the centre, so position the part under it and set the gap with the knurled nut.\u003c\/p\u003e\n\u003cp\u003e\u003csmall\u003eAnswered by John Legg, Director, Debug Store\u003c\/small\u003e\u003c\/p\u003e\n\n\u003cp\u003e\u003csmall\u003eLast reviewed: 18 November 2026\u003c\/small\u003e\u003c\/p\u003e\n\u003cdetails\u003e\u003csummary\u003e\u003csmall\u003eSources (2 manufacturer pages)\u003c\/small\u003e\u003c\/summary\u003e\u003cul\u003e\n  \u003cli\u003e\u003ca href=\"https:\/\/www.omicron-lab.com\/products\/vector-network-analysis\/accessories\/impedance-test-fixtures-b-wic-b-smc\"\u003eOMICRON Lab Impedance Test Fixtures B-WIC and B-SMC\u003c\/a\u003e\u003c\/li\u003e\n  \u003cli\u003e\u003ca href=\"https:\/\/us-shop.omicron-lab.com\/products\/b-smc-impedance-test-fixture-for-surface-mounted-components\"\u003eOMICRON Lab Shop: B-SMC Impedance Test Fixture\u003c\/a\u003e\u003c\/li\u003e\n\u003c\/ul\u003e\u003c\/details\u003e","brand":"OMICRON-Lab","offers":[{"title":"Default Title","offer_id":42541401276639,"sku":"P0005759","price":650.0,"currency_code":"GBP","in_stock":true}],"thumbnail_url":"\/\/cdn.shopify.com\/s\/files\/1\/0522\/6931\/8333\/products\/omicron-lab-impedance-adapter-omicron-lab-b-smc-impedance-adapter-for-sm-components-36889492848863.jpg?v=1684970931","url":"https:\/\/thedebugstore.com\/it\/products\/omicron-lab-b-smc-impedance-adapter-uk","provider":"Debug Store","version":"1.0","type":"link"}